The Baseline® 9000 H Heated Hydrocarbon Analyzer is a microprocessor based instrument designed for continuous ambient or process hydrocarbon gas measurement in environmental or industrial settings. The heated total hydrocarbon analyzer is configured for single point analysis (with or without a sample pump) of samples heated up to 376 °F (191 °C) for pre-filtered (< 10 microns) non-condensing samples.
Detection limits down to < 0.1 ppm. User-programmable ranges from 10 ppm to 10% (propane) or 10 ppm to 50% (methane) are factory-configured per the customer’s application to facilitate installation and setup.
Using a Flame Ionization Detector (FID), AMETEK MOCON Baseline® FlowGuard electronic control regulates the delivery of fuel, air, and a small portion of the sample gas, to the FID. During the combustion process, organic or hydrocarbon-based gases in the sample are ionized, detected by the instrument, and then reported as a concentration. The automatic calibration feature enhances the long-term analytical stability of the
instrument.
All instrument parameters are reported clearly and continually refreshed on a large, graphical LCD display. Using analog, digital, and logic output communication capabilities, analytical information from the analyzer can be acquired using an external PC and a simple communications program such as Windows® HyperTerminal or the analyzer can output binary or ASCII formats directly to a data acquisition system or PLC. Every Baseline® 9000 H analyzer includes a AMETEK MOCON's free PC utility 9000 Keeper used for storing and uploading multiple methods, as well as sending configuration settings, directly to the analyzer.
Automated Control Features
- FlowGuard control of fuel, air, and sample with automatic shut-off
- Electronic back-pressure regulator with sample bypass system
- Automatic FID ignition
- Automatic calibration at user-defined intervals
Versatile Platform
- Graphic LCD display with easy to use menu
- Benchtop or rack-mountable
- Ethernet or serial customizable output
- Programmable analog output ranges
- Programmable relays for diagnostics, concentration, alarms, and events
Baseline® 9000 Heated Hydrocarbon Analyzer
|
Units
|
Detector
|
Flame Ionization (FID)
|
Ranges
|
User definable based upon calibration (as methane):
- Very Low – 0.15 ppm to 200 ppm
- Low – 0.3 ppm to 2,000ppm
- Medium – 0.6 ppm to 20,000 ppm
- High – 0.003% to 50%
|
Accuracy, Repeatability
|
±1% full-scale
|
Drift, Zero
|
±0.025% full-scale, 24 hours
|
Drift, Span
|
±1% full-scale, 24 hours
|
Response Time
|
T90 < 5 seconds
|
Sampling
|
Internal single point, pump optional
|
Alarms
|
Multilevel concentration and fault alarms, audible and visual
|
Calibration
|
Programmable automatic or manual calibration
|
Support Gas Requirement
|
UHP Hydrogen (H2) – 30 cc/min
Zero Air – 175 cc/min
Span gas – methane is typical
Fuel blend options available
|
Display
|
Graphical LCD, 3.4” x 4.5” (8.64 cm x 11.43 cm)
|
Outputs, Standard
|
RS-232
LAN/Ethernet
Analog, 1 programmable 0-20mA/4-20mA isolated output
Relays, 5 programmable Form A relays rated to 3A @230VAC
|
Operating Temperature
|
32 to 104°F (0 to 40°C)
|
Operating Humidity
|
0 to 95% (non-condensing)
|
Configuration
|
Benchtop or 19” (48.3cm) rack-mount, 3U
|
Other Specifications (all models):
Physical Specifications
- 5.25” H x 19” W x 16.25” D (13.02 cm x 48.26 cm x 41.28 cm)
- <20 Lbs. (9.07 kg)
Electrical Requirements
- Voltage: 100 - 240 VAC 50/60 Hz, 1 A
Preventive Maintenance Kit
- Includes all parts potentially needed on-hand
Gas Generators
- Hydrogen generators
- Zero Air generators
Software
- 9000 Keeper provided for method switching
Service & Support Options
- Start-up and training, 1 day onsite
Environmental Monitoring
- AMC in Cleanrooms
- Air Quality Monitoring
- Fugitive Emission Monitoring
Industrial Process Analysis & Control
- Beverage Gas Monitoring
- Industrial Gas Mixing Control
- Process Gas Analysis
- Specialty and Industrial Gas Monitoring
- Surface Logging Analysis